Download PDFOpen PDF in browserVariability in Coronal Knee Laxity Measured During Computer-Assisted Total Knee Arthroplasty4 pages•Published: September 25, 2020AbstractPrincipal component analysis on 376 TKA knees presented the distinctive patterns and variabilities in the coronal knee angular laxity throughout the range of motion, measured post-implantation during computer-assisted total knee arthroplasty. The variability in the laxity curves were dominated by the first mode of variation (varus/valgus offset of the laxity envelope) and the second mode of variation (varus/valgus crossing pattern). Further analysis revealed that surgeon-specific impact was associated with the first mode of variation for the laxity curve, and both first and second mode of variation for the size of the laxity envelope. The results shed a light on the characteristics and variabilities of post-implantation soft-tissue laxity under surgical reality and may be used to further understand the clinical implications of intraoperative soft-tissue management.Keyphrases: intraoperative laxity curve, principle component analysis, total knee arthroplasty, variability In: Ferdinando Rodriguez Y Baena and Fabio Tatti (editors). CAOS 2020. The 20th Annual Meeting of the International Society for Computer Assisted Orthopaedic Surgery, vol 4, pages 45-48.
|